A   B   C   D   E   F   G
H   I   J   K   L   M   N
O   P   Q   R   S   T   U
V   W   X   Y   Z

FINE PARTICULATE SAMPLER

Fine Particulate Sampler

SPECTRO’s model SLE-FPS105 is an advanced fine dust sampler, which can be used to determine the concentration of particulate matter of size less than 2.5μ in atmosphere. It has been specifically designed to comply with EPA reference method for determination of PM2.5, when used without Impactor (2.5μ cut stage), the system may be optionally used for particulate matter less than 10μ.

Ambient air enters the SLE-FPS105 system at a constant sampling rate at 1 m3/hr by a suitable critical orifice through an omni-directional inlet designed to provide a clean aerodynamic cut-point for particles greater than 10 microns. Particles in the air stream finer than 10 microns proceed to a second impactor that has an aerodynamic cut point at 2.5 microns. The air sample and fine particulates exiting from the PM2.5 impactor is passed through a 47mm diameter Teflon filter membrane that retains the FPM. Instrument has inbuilt oil free brushless blower and dry gas meter to have direct volume of air passed through the system.

Specifications

Model : SLE-FPS105
Particle Size
 
 
 
:Omni-directional air inlet with PM 10 separation through an impactor followed by PM 2.5 separation through a WINS Impactor
Sampling rate
 
 
 
:Constant sampling rate of 1m3/hr unaffected by voltage fluctuation and filter choking maintained by critical orifice system
Filter Media
 
: Filter holder designed to hold any standard 47 mm dia filter
Sample Volume : Dry Gas meter records the total air volume sampled
Power Req. : Single phase AC 220 Volts, 50 Hz

Features

  • PM10 and PM2.5 Impactors of sampler based on designs standardized by US EPA
  • Same instrument can be used for PM10 and PM2.5 sampling
  • Lower constant sampling rate of 1m3/hr
  • Compact and portable
  • Sturdy construction for outdoor use
  • Easy setup and operation


  • Back to Environment Equipments